This paper describes a motion analysis ultrasound system (MAUS) designed for the in-vivo study of the kinematics of patellofemoral joints (PFJ) in both normal and replaced knees. This system utilises non-iodising radiation to effectively acquire kinematic data during weight-bearing activities. Validation studies on a phantom established that the measurement accuracy of the system was 1.43 mm. A clinical validation trial is included. © 2013 IEEE.

Original publication

DOI

10.1109/MeMeA.2013.6549747

Type

Journal article

Journal

MeMeA 2013 - IEEE International Symposium on Medical Measurements and Applications, Proceedings

Publication Date

15/08/2013

Pages